07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 41
Detailed Instructions
Programming SINGLE DOSE Mode
SINGLE DOSE mode delivers a programmed volume of drug over a specified period of
time. The infusion is programmed in terms of dose volume and dose duration.
1. At the SELECT MODE ▲▼ prompt, use ▲ or ▼ to display the available infusion mode
selections. Press
CONFIRM to select the desired infusion mode.
2. Use ▲ or ▼ and CONFIRM as required to select the syringe manufacturer and size.
3. Enter the dose volume (INFUSE) in mL. Press CONFIRM to complete the data entry.
4. Enter the dose duration (OVER) in hours and minutes. Press CONFIRM to complete
the data entry.