Configuration
Page 64 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
For each enabled manufacturer, the pump displays a VIEW (name) prompt, where (name)
is the manufacturers name. A pending
Y response is also displayed. To begin reviewing
the enabled syringe sizes for that manufacturer, press
CONFIRM to accept the Y
response.
Syringe size descriptions are displayed in the text field. Press
CONFIRM to view each
additional syringe.
To skip a manufacturer, press either or to change the
Y to N and press CONFIRM.
The available manufacturer names and sizes are listed in the “Syringe Data Tables” on
page 25.
Defaults Group (VIEW DFAULTS)
This option group displays the default options that are enabled. The pump allows four
default options: Default Library (
DFAULT LIBRARY), Default Protocol (DFAULT
PROTOCOL
), Default Mode (DFAULT MODE) and Default Manufacturer (DFAULT MFR).
The defaults enable the pump to recall the selection that was used for the previous
infusion, and to provide that selection in response to the first or keypress. This
feature is especially useful when a particular protocol library, protocol, infusion mode, or
syringe manufacturer is likely to be used repeatedly.
Miscellaneous Group (VIEW MISC)
This displays the miscellaneous options that are enabled. The Miscellaneous group
options are: Remote Control, Protocol Library, Size Override, Occlusion Sensitivity, Audio
Range, Rate Range, Auto Lock, Backlight, and Enable DB9 Port.
Remote Control
This option allows the operator to control the pump from a remote device, such as a
computer. A standard 9-pin connector (DB9) located on the rear of the pump is provided
for an RS-232(C) communications link between the pump and the computer. See the
“Auto Syringe AS50 Infusion Pump Programmer’s Manual” for details.
Protocol Library (LIBRARY)
This option allows the operator to select from one or more protocol libraries containing
frequently used program data. Protocols are defined by authorized personnel according
to procedures detailed in the “Auto Syringe AS50 Infusion Pump Service Manual.”
Size Override
SIZE OVERRIDE modifies the syringe programming process by allowing the operator to
select a syringe size that differs from the size detected by the pump. Instead of a
VERIFY
xx ML prompt, the pump displays SELECT SIZE . When or is pressed, the pump
displays the syringe size that best matches the syringe. However, the operator can press
or to select a different size. When
CONFIRM is pressed to complete the syringe
selection, if the syringe size selected appears different from the detected syringe size, the
pump displays a
VERIFY AGAIN prompt and the operator must press CONFIRM again to
complete the selection.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals