07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 83
Alerts and Alarms
Failsafe Alarms
Failsafe alarms occur only when proper operation cannot be assured. The infusion cannot
be restarted, and must be reprogrammed after the problem has been corrected. The
pump must be turned off to clear the alarm.
Message Cause Correction
(ERROR CODE)
(continuous two-tone audio,
SYSTEM light on)
An internal error has been detected. Repair is required. Make a note of
all displayed error codes and the
conditions prior to the error. Take
the pump out of service.
Note: In some cases, there may not
be a legible error code display.
CHARGE DEPLETED
(continuous two-tone audio,
SYSTEM and BATTERY lights on)
The battery voltage is below
minimum operating level.
Connect the charger and recharge
the battery for a minimum of 16
hours.
FUSE FAILURE (continuous
two-tone audio, SYSTEM and
BATTERY lights on)
Fuse self-test has detected a fuse
failure.
Repair is required. Take the pump
out of service.
POWER FAULT (continuous,
two-tone long beeps, SYSTEM and
BATTERY lights on)
The pump’s internal power supply is
out of tolerance, or there is a fault in
the charger.
Repair is required. Take the pump
and charger out of service.