Detailed Instructions
Page 28 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Preliminary Operations
Manual Purge
1. Attach a Baxter 2N Series or equivalent IV extension set to the filled syringe.
2. Remove the sterile cap from the end of the tubing.
3. Manually purge until a drop of fluid is visible at the end of the tubing.
4. Replace the sterile cap on the end of the tubing.
Mount the Syringe
1. Grasp the FINGER GRIP (1) on the PLUNGER DRIVER (2) and pull it all the way out.
2. Slide the PLUNGER DRIVER to the top of the pump and release the FINGER GRIP.
2
1
7