EasyManuals Logo

baxter Auto Syringe AS50 User Manual

baxter Auto Syringe AS50
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
Detailed Instructions
Page 28 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Preliminary Operations
Manual Purge
1. Attach a Baxter 2N Series or equivalent IV extension set to the filled syringe.
2. Remove the sterile cap from the end of the tubing.
3. Manually purge until a drop of fluid is visible at the end of the tubing.
4. Replace the sterile cap on the end of the tubing.
Mount the Syringe
1. Grasp the FINGER GRIP (1) on the PLUNGER DRIVER (2) and pull it all the way out.
2. Slide the PLUNGER DRIVER to the top of the pump and release the FINGER GRIP.
2
1
7

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals