07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 51
Detailed Instructions
Number Out of Range
Every programmable field is limited to a particular range of allowed program
values, depending on the syringe, drug concentration, and other factors.
Attempting to program a value outside the acceptable range results in a
PUMP
LIMIT
alert. The pump beeps and substitutes the nearest acceptable value.
Press
CONFIRM to accept the value, or program a new value.
Data Test
Whenever a data field is edited, the pump tests the remaining program data
fields for range limits. If there is a conflict, the affected fields are cleared to
dashes, and the first cleared field is automatically opened for editing. As each
cleared field is reprogrammed, the next cleared field (if any) is automatically
opened. All cleared fields must be reprogrammed before the infusion can be
started.
Example: Reprogramming the SYRINGE MFR field automatically clears the
SYRINGE SIZE field.