07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 11
Simplified Instructions
Programming
Note: If the Protocol Library feature is enabled, a ▲▼ SELECT LIBRARY prompt is dis-
played first. Use ▲, ▼ and
CONFIRM as required to select a protocol library. Next,
select a protocol from the library.
If the selected protocol specifies a default infusion mode, the pump skips step 1.
1. At the SELECT MODE ▲▼ prompt, use ▲ or ▼ as necessary to display the available
infusion mode selections. Press
CONFIRM to select an infusion mode.
2. Select the syringe manufacturer and size using ▲ or ▼ as necessary and complete
the data entry by pressing
CONFIRM. (For use of the Size Override feature, see “Size
Override” on page 64.)
3. Enter the remaining program data as required. The current field is the one that is
flashing.
• Press number keys to enter the desired value.
• Press CONFIRM to complete the data entry.
4. To change programmed data:
• Press ▲ or ▼ to move to the desired field.
• Press EDIT. The current value flashes.
• Press ▲, ▼, or number keys to change the value.
• Press CONFIRM to complete the data entry.
5. To program or edit the bolus size:
• Press BOLUS, EDIT.
• Set the bolus size using the number keys.
• Press START to deliver a bolus, or CONFIRM to store the bolus size.
6. To clear the Total Delivered display:
• Use ▲ or ▼ to move to the TOTAL mL field.
• Press EDIT, CLR to reset the total to 0.00.