07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 27
Detailed InstructionsCha pter 1
This section provides detailed descriptions of the seven basic steps required to operate
the pump:
Note: Steps 1 and 2 can be performed after programming the pump if necessary (for
example, if using CUSTOM DILUTION mode).
1. Attach a Baxter 2N series (or equivalent) extension set to the syringe and manually
purge the air from the filled syringe and tubing.
2. Mount the syringe on the pump.
3. Slide the ON/OFF switch (7) to the “on” position (refer to figure on the following page).
4. Program the pump.
5. Purge the syringe and tubing again to remove slack from the plunger driver.
6. Connect the extension set to the patient.
7. Press START to begin the infusion.