Configuration
Page 66 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
Idle Alarm
This option enables the pump to issue a continuous audible alarm when a PUMP IS IDLE
condition occurs. If the option is disabled, the pump beeps 15 times at two minute
intervals when a
PUMP IS IDLE condition occurs.
Near End Alert (NEAR END)
This option enables the pump to issue an alert during an infusion prior to an expected
VOL LIMIT or EMPTY alarm. The option specifies the number of minutes prior to the alarm
that the expected alert will be issued and the number of beeps in the alert.
The form of a Near End alert display is xx
MIN VOL LIM, xxMIN EMPTY or xxMIN (protocol).
The value xx is the configured number of minutes. The first and second forms apply to
impending
VOLUME LIMIT or EMPTY conditions when no protocol has been selected. The
third form applies to either condition when a protocol has been selected. Press
CONFIRM
to view the current settings.
Note: When an infusion is started, no
NEAR END alert will occur if fewer than the config-
ured number of minutes remain before an
EMPTY or VOL LIMIT alarm is due to
occur. For example, if the
NEAR END alert is configured for 10 minutes and 9 min-
utes remains before the volume limit is reached, no
NEAR END alert will occur
when
START is pressed.
Bolus Complete Alert (BOLUS COMPLETE)
This option enables the pump to issue an alert when a bolus completes.
Dose Complete Alarm (DOSE COMPLETE)
This option enables the pump to issue an alarm when a dose completes in SINGLE
DOSE mode.
Volume Limit Alarm (VOLUME LIMIT)
This option applies to mL/hr and mL/min modes only. During initial programming, the
pump prompts the operator to enter a volume limit value in mL. The Volume Limit feature
triggers an alarm after the pump has delivered the pre-programmed volume of drug. With
this option enabled, a
NEAR END alert is issued when the Volume Limit will be reached
within a preconfigured time period at the current rate of infusion. See “Near End Alert” on
page 66.
Identifier
The Identifier is not a configuration group, but is included here for reference purposes.
The Identifier is a short message (up to eight characters) that is displayed for three
seconds following the Lamp Test. The Identifier can be created or edited during the pump
configuration process.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals