07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 5
PrecautionsC hapter 3
Federal (USA) law restricts this device to sale by or on the order of a physician.
The charger port is to be used only with “900 Series” chargers or other accessories that are
labeled specifically for use with Auto Syringe AS40A or AS50 Series Infusion Pumps, including
the 1M8563 Multi-Charger.
CAUTION: Refer all service, repair, and calibration to qualified technical
personnel.
CAUTION: Do not operate the Auto Syringe AS50 Infusion Pump in the
presence of flammable anesthetics, oxygen-enriched, or explosive
atmospheres.
● Though the factory-supplied configuration settings are suitable for most therapies, the
operator and hospital professionals should verify that the pump’s settings are
appropriate for the clinical application.
● The Protocol Library feature is a configuration option developed to facilitate infusion
programming. Before using a protocol for any specific drug, refer to the full
prescribing information supplied by the drug manufacturer.
● As with all medical electronic equipment, care must be exercised to avoid exposing
this device to powerful sources of electromagnetic interference.
● Do not expose the pump to X-rays, gamma rays, or other ionizing radiation, or to
strong electric or magnetic fields.
● Do not autoclave, steam sterilize, EtO sterilize, or subject the pump or charger to
temperatures in excess of 104° F (40° C).
● Do not use the following chemicals on the device, as they will damage the front panel:
acetone, acetoaldehyde ammonia, benzene, hydroxytoluene, methylene chloride,
and ozone, Do not use cleaners containing n-alkyl dimethyl ethylbenzyl ammonium
chloride unless they appear in the list of recommended cleaners in the Pump
Maintenance section of this manual.
● Do not use hard or sharp objects on the keypad.
!
!
!
!