07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 23
General Information
Infusion Modes
The pump provides 18 infusion modes, which fall into two major categories: continuous
and timed infusions. Continuous infusions deliver at a steady, programmed rate. Timed
infusions deliver one or more equal doses according to a programmed schedule. The
infusion modes are briefly described below. For more information see “Detailed
Instructions” on page 27.
Continuous Infusion Modes
mL/hr and mL/min: The rate of infusion is programmed in mL/hr or
mL/min, depending on the mode selected. A volume limit
can be programmed for these modes only.
Units/hr, Units/min, mUnits/hr and mUnits/min: The drug concentration is entered in
units/mL, and the dose is entered in Units/hr, Units/min,
mUnits/hr or mUnits/min, depending on the mode selected.
The pump automatically displays the effective rate in mL/hr.
mg/hr, Mg/min, mcg/hr and mcg/Min: The drug concentration is entered in mg/mL, and
the dose is entered in mg/hr, mg/min, mcg/hr or
mcg/min, depending on the mode selected. The pump
automatically displays the effective rate in mL/hr.
mg/kg/hr, mg/kg/min, mcg/kg/hr and mcg/kg/min: The patient's body weight is
entered in kg, the drug concentration is entered in mg/mL,
and the dose is entered in mg/kg/hr, mg/kg/min, mcg/kg/hr
or
mcg/kg/min, depending on the mode selected. The pump
automatically displays the effective rate in mL/hr.
CUSTOM DILUTION: In this mode the pump is programmed for a continuous
infusion and the drug concentration is adjusted for a target
infusion rate and dose. The target rate is entered in mL/hr,
the patient’s body weight is entered in kg, the initial drug
concentration is entered in mg/mL, the target dose is
entered in mcg/kg/min, and the final volume is entered in
mL. The pump calculates and displays the volume of drug
(in mL) to be added to the infusion base solution to produce
a final concentration that achieves the target rate and target
dose. The final concentration is displayed when the infusion
is initiated.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals