07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 31
Detailed Instructions
Check the Lamp Test
As soon as the pump is turned on, an automatic self-test
begins. This includes the Lamp Test, which tests the status
display and the status panel lights. Verify that all the lights
illuminate and that all parts of the status display are legible.
Carefully check the Lamp Test daily, or once per shift. See
“Display, Battery, and Syringe Sensor Check” on page 75.
To examine the status display more closely during the
Lamp Test, press and hold
STOP while the status display
characters are all visible. The Lamp Test will complete
when
STOP is released.
If
CONFIRM is pressed during the Lamp Test, the battery
voltage will be displayed in the text field immediately
following completion of the Lamp Test.
The Lamp Test ends with a short beep. If the pump is
configured to show an identifier message, it will appear in
the text field at the bottom of the status display.
This completes the preliminary operations.