07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 65
Configuration
Note: To restart the syringe selection process, press CLR before pressing CONFIRM to
complete syringe selection.
Occlusion Sensitivity (PSI RANGE)
This option sets the occlusion detection system operating range. Three choices are
available:
H (Highest back pressure), M (Medium setting), L (Lowest back pressure). For
more details see the “Auto Syringe AS50 Infusion Pump Technical Manual.”
Audio Range
The audio volume level can be configured for: H (High) or L (Low). The recommended
setting is
H. The quieter L setting should only be used in noise-sensitive environments
where the infusion is constantly monitored.
Rate Range
RATE RANGE can be configured for H (High, up to 438 mL/hr), M (Medium, 120 mL/hr) or
L (Low, 15 mL/hr). The maximum delivery rate will be limited to the lesser of the Rate
Range option setting or the syringe maximum (see the “Syringe Data Tables” on
page 25).
The minimum delivery rate is as shown in the table, regardless of the Rate Range setting.
Auto Lock
If an infusion has been programmed and the pump is in the Run state, the Auto Lock
feature automatically turns on the Lock function two minutes after the last keypress.
Backlight (VIEW BACKLITE)
The status display backlight is lit whenever the charger is connected and the pump is
powered on. The
VIEW BACKLITE option turns the backlight on after a keypress when the
charger is not connected. The backlight turns off automatically after a preset time period
elapses (0 to 60 minutes). Press
CONFIRM to view the current setting.
Enable DB9 Port
When this option is set to Y (yes), the standard 9-pin connector (DB9) located on the rear
of the pump is enabled as the serial RS-232(C) communication link (to allow remote
control operation). When this option is set to
N (no), the RS-232(C) serial communication
link is disabled.
Alerts Group (VIEW ALERTS)
This option group displays the alert and alarm options that are enabled. The Alerts group
options are:
IDLE ALARM, NEAR END ALERT, BOLUS COMPLETE ALERT, DOSE
COMPLETE ALARM
, and VOLUME LIMIT ALARM.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals