07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 81
Alerts and Alarms
Alarms
Message Cause Correction
CHECK BARREL
(continuous beeps and flashing
ALERT light)
During motor movement the pump
has detected that the barrel clamp is
not fully closed on the syringe
barrel, or the syringe barrel is not
properly seated in the cradle.
Pressing any key silences the audio
portion of the alarm. Make sure the
syringe is centered in the barrel
clamp cradle, and that the syringe
flange is in the barrel clamp flange
slot. To ensure proper seating, use
the heel of the hand to press the
barrel clamp firmly against the
syringe. Motor movement is not
allowed until the syringe is properly
mounted.
CHECK FLANGE
(continuous beeps and flashing
ALERT light)
During motor movement the pump
has detected that the syringe flange
is not seated in the flange slot.
Pressing any key silences the audio
portion of the alarm. Remove the
syringe and remount it. Make sure
the syringe is centered in the barrel
clamp cradle, and that the syringe
flange is in the barrel clamp flange
slot. To ensure proper seating, use
the heel of the hand to press the
barrel clamp firmly against the
syringe. Motor movement is not
allowed until the syringe is properly
mounted.
CHECK PLUNGER
(continuous beeps and flashing
ALERT light)
During motor movement the pump
has detected that the plunger
capture mechanism has not
captured the syringe plunger.
Pressing any key silences the audio
portion of the alarm. Grasp the
finger grip on the plunger driver and
pull it all the way out. Slide the driver
down until it contacts the syringe
plunger. Push in firmly. Motor
movement is not allowed until the
syringe is properly mounted.
CHECK SYRINGE
(continuous beeps and flashing
ALERT light)
At least two of the three syringe
mounting points (flange, barrel,
plunger) are not correctly
positioned.
Pressing any key silences the audio
portion of the alarm. Check flange,
barrel, and plunger, and ensure that
they are mounted correctly. Motor
movement is not allowed until the
syringe is properly mounted.
DOSE COMPLETE
(continuous beeps and flashing
ALERT light)
(Single Dose Infusion Mode only)
The programmed dose has been
delivered.
Pressing any key silences the
alarm.
DOSE DUE
(continuous beeps and flashing
ALERT light)
(Manual Schedule Infusion Mode
only) The NEXT DOSE IN timer has
reached 0:00, and the next dose is
due to be delivered.
Pressing any key silences the audio
portion of the alarm. Press START
to deliver the scheduled dose.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals