Alerts and Alarms
Page 82 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
EMPTY or
EMPTY (protocol)
(continuous beeps and flashing
ALERT light)
The syringe is empty. Pressing any key silences the audio
portion of the alarm. Release and
slide the plunger driver back at least
1.6 inches (4 cm) to reset the alarm.
Note: due to variability in syringe
dimensions, a LINE OCCLUDED
alarm may occur before the EMPTY
alarm. Option: The infusion may be
restarted without resetting EMPTY
condition. The pump will continue to
display EMPTY and ALERT light will
continue flashing. The infusion will
continue until STOP is pressed or a
LINE OCCLUDED alarm occurs.
FINISH EDITING
(continuous beeps and flashing
ALERT light)
The Rate or Dose field is being
edited while an infusion is in
progress, and there has been no
keypress for 15 seconds.
Pressing any key silences the audio
portion of the alarm. Complete the
editing operation.
LINE OCCLUDED
(continuous beeps and flashing
ALERT light)
There is an occlusion, line
constriction, or other condition that
causes excess plunger force.
The alarm automatically stops the
pump. Pressing any key silences the
audio portion of the alarm. Check
for empty syringe, kinked tubing,
clogged catheter, etc. Check for
foreign material preventing
movement of the pump mechanism.
RELIEVE RESIDUAL SYRINGE
PRESSURE BY RELEASING THE
PLUNGER DRIVER. If the plunger
driver is not released, an
unintentional bolus may occur when
the blockage is cleared. The
message clears when the infusion is
successfully restarted.
PUMP IS IDLE
(continuous beeps and flashing
ALERT light)
The pump has been in Standby
state for at least 2 minutes. Also
occurs if there is no keypress within
15 seconds after the infusion is
stopped.
Pressing any key silences the audio
portion of the alarm. The Lock
function suppresses the audio
portion of this alarm.
Note: If the keypad is locked, there
will be no beeps, but the PUMP IS
IDLE message still appears.
VOLUME LIMIT
(continuous beeps and flashing
ALERT light)
The programmed volume limit has
been reached.
Pressing any key silences the audio
portion of the alarm. Restarting the
pump (by pressing START or
initiating a bolus) resets the Volume
Limit field.
Message Cause Correction

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals