07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 37
Detailed Instructions
Bolus in mg/hr, mg/min, mcg/hr, and mcg/min Modes
Note: If a protocol has been selected, bolus delivery may be disallowed or a default
bolus size may appear automatically. If a default bolus size is displayed, use
and or the digit keys to modify the value as necessary.
In these modes, bolus size is entered in mg or mcg. The pump automatically calculates
the equivalent bolus size in mL and displays both values. Bolus rate is displayed in mL/hr.
See “Bolus Function” on page 55 for more information on bolus delivery.
Programming Notes
1. The drug concentration range is from 0.0001 mg/mL to 100.0 mg/mL. The dose limits
are determined by the drug concentration and syringe selection. See “Syringe Data
Tables” on page 25 for minimum and maximum infusion rates for each syringe.
2. The CONC. field cannot be reprogrammed once the infusion has been started.
3. Pressing CONFIRM during delivery briefly displays additional program data.
4. A bolus can only be programmed, edited, or delivered after the drug concentration
(
CONC.) has been programmed.
Programming mg/kg/hr, mg/kg/min, mcg/kg/hr, and mcg/kg/min Modes
In these modes, the pump runs at a programmed, constant rate. The infusion is
programmed in terms of patient body weight, drug concentration and dose.
Note: If a protocol has been selected, the infusion mode may have been selected auto-
matically.
1. At the SELECT MODE ▲▼ prompt, use or to display the available infusion mode
selections. Press
CONFIRM to select the desired infusion mode.
2. Use or and CONFIRM as required to select the syringe manufacturer and size.
3. Enter the patient body weight (BODY WT) in kg. Press CONFIRM to complete the data
entry.
Note: If a protocol has been selected, a default drug concentration may appear automat-
ically. If a default drug concentration is displayed, use
and or the digit keys to
modify this value as necessary, and press
CONFIRM to complete the data entry.
4. Enter the drug concentration (CONC.) in mg/mL. Press CONFIRM to complete the data
entry.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals