07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 61
ConfigurationChapter 1
The pump can be custom-configured to meet particular needs, or to simplify infusion
programming. Instructions for setting configuration options are given in the Auto Syringe
AS50 Infusion Pump Technical Manual.
Configuration Review
The Configuration Review feature allows the pump’s configuration to be examined without
risk of accidental alteration. This feature is activated by entering the code number
123 at
the initial prompt (
SELECT MODE ▲▼ or SELECT PROTOCOL ▲▼). The error beeps that
occur when
123 is entered are provided to discourage patient tampering and may be
ignored.
Date and Time Display
After 123 is entered, the date and time are briefly displayed. The time is displayed in hours
and minutes, separated by a decimal point (
HH.MM).
Software and Configuration Version Display
Following the time and date display, the software and configuration version numbers are
briefly displayed.
The first line of the text display indicates the pump software version. The second line is
divided into two fields. The left field displays the factory configuration version, and the
right field displays a four character checksum that uniquely identifies the current
configuration (including protocols), as shown below.
The checksum field allows a high-confidence rapid comparison of configurations between
two pumps.
Reviewing Configuration Options
The configurable options are organized into several groups, each of which begins with a
prompt that requires a “yes” or “no” response. Each option group is identified by name in
the text display field (for example:
VIEW MODES). The syringe manufacturer field displays
Y (for “yes”), which represents the pending response.
Example: The display for software version 6.00, factory configuration version D1, and a current
configuration checksum of 3AC5, would be shown on the display as follows:
software version
V 7.0 0
factory configuration D 1 3 A C 5 configuration checksum

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals