07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 13
Simplified Instructions
Programming the Timed Infusion Modes
Programming for each infusion mode proceeds in the sequence shown in the following
table, in top-to-bottom order.
Timed Infusion Modes
Steps SINGLE DOSE MANUAL SCHEDULE AUTO SCHEDULE
1 Select mode, syringe
manufacturer, syringe size.
Select mode, syringe
manufacturer, syringe size.
Select mode, syringe
manufacturer, syringe size.
2
INFUSE field: Enter dose in mL. INFUSE field: Enter dose in mL. INFUSE field: Enter dose in mL.
3
OVER field: Enter dose duration
in hours and minutes (hh:mm).
OVER field: Enter dose duration
in hours and minutes (hh:mm).
OVER field: Enter dose duration
in hours and minutes (hh:mm).
4
EVERY field: Enter dose Interval
in hours and minutes (hh:mm).
EVERY field: Enter dose Interval
in hours and minutes (hh:mm).
5
NEXT DOSE IN field: Enter delay
to start of first dose in hours and
minutes (hh:mm).
To start the first dose without delay,
program
NEXT DOSE IN field to
0:00.
NEXT DOSE IN field: Enter
delay to start of first dose in hours
and minutes (hh:mm).
To start the first dose without
delay, program
NEXT DOSE IN
field to
0:00.
6 When the
DOSE DUE alarm
occurs, press START to deliver
one dose.
Note:
• If the Protocol Library feature is active, these modes can be accessed directly by selecting
NONE (if available) at the
SELECT LIBRARY or SELECT PROTOCOL programming step.