07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 25
General Information
Syringe Data Tables
This table shows the performance ranges for infusion delivery parameters, which are a
function of the syringe dimensions.
Infusion Rate Ranges
Syringe
Manufacturer
Syringe
Size
(mL)
Minimum
Flow
Rate
(mL/hr)
Maximum
Flow
Rate
(mL/hr)
Minimum
Flow Rate
(mL/min)
Maximum
Flow Rate
(mL/min)
B-D Plastic 1
3
5
10
20
30
60
0.01
0.02
0.03
0.1
0.1
0.1
0.1
11
38
74
107
188
241
366
0.001
0.001
0.001
0.002
0.002
0.002
0.002
0.183
0.633
1.233
1.783
3.13
4.01
6.1
Monoject 1
3
6
12
20
35
60
0.01
0.02
0.03
0.1
0.1
0.1
0.1
11
41
83
126
208
285
366
0.001
0.001
0.001
0.002
0.002
0.002
0.002
0.183
0.683
1.383
2.1
3.46
4.75
6.1
Te ru mo 1
3
5
10
20
30
60
0.01
0.02
0.03
0.1
0.1
0.1
0.2
11
41
87
129
210
276
438
0.001
0.001
0.001
0.002
0.002
0.002
0.004
0.183
0.683
1.45
2.15
3.5
4.6
7.3
Maximum Flow Rates shown apply to pumps configured for:
Rate Range = H (High).
Rate Range = High limits maximum rate to maximum rate for syringe used.
Rate Range = Medium limits maximum rate to 120 mL/hr (2 mL/min).