07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 25
General Information
Syringe Data Tables
This table shows the performance ranges for infusion delivery parameters, which are a
function of the syringe dimensions.
Infusion Rate Ranges
Syringe
Manufacturer
Syringe
Size
(mL)
Minimum
Flow
Rate
(mL/hr)
Maximum
Flow
Rate
(mL/hr)
Minimum
Flow Rate
(mL/min)
Maximum
Flow Rate
(mL/min)
B-D Plastic 1
3
5
10
20
30
60
0.01
0.02
0.03
0.1
0.1
0.1
0.1
11
38
74
107
188
241
366
0.001
0.001
0.001
0.002
0.002
0.002
0.002
0.183
0.633
1.233
1.783
3.13
4.01
6.1
Monoject 1
3
6
12
20
35
60
0.01
0.02
0.03
0.1
0.1
0.1
0.1
11
41
83
126
208
285
366
0.001
0.001
0.001
0.002
0.002
0.002
0.002
0.183
0.683
1.383
2.1
3.46
4.75
6.1
Te ru mo 1
3
5
10
20
30
60
0.01
0.02
0.03
0.1
0.1
0.1
0.2
11
41
87
129
210
276
438
0.001
0.001
0.001
0.002
0.002
0.002
0.004
0.183
0.683
1.45
2.15
3.5
4.6
7.3
Maximum Flow Rates shown apply to pumps configured for:
Rate Range = H (High).
Rate Range = High limits maximum rate to maximum rate for syringe used.
Rate Range = Medium limits maximum rate to 120 mL/hr (2 mL/min).
Rate Range = Low limits maximum rate to 15 mL/hr (0.25 mL/min).

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals