07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 67
Configuration
Terminating Configuration Review
When the last configuration group option has been reviewed, the pump automatically
restarts the Lamp Test, as if it had just been turned on.
To cancel Configuration Review at any time, turn the pump off.
Custom Configuration
The pump can be reconfigured as often as desired, to select features that best suit the
needs of the healthcare facility.
If there is any question regarding the pump’s current configuration or applicability for a
particular clinical application, the operator and healthcare professionals should verify that
the settings are appropriate.
While no special tools are needed, the configuration should be changed only by
authorized personnel.
Factory Standard Configuration
The factory standard configuration is:
Protocol Library: See the chart on the following page.
Modes: ML/HR MCG/KG/MIN
SINGLE DOSE MCG/KG/HR
MCG/MIN
Manufacturers and Sizes: B-D 1, 3, 5, 10, 20, 30, 60 mL Plastic
Monoject 1, 3, 6, 12, 20, 35, 60 mL
Terumo 1,3,5,10, 20, 30, 60 mL
Defaults: All enabled
Misc: PSI RANGE = M (Medium)
AUDIO RANGE = H (High)
RATE RANGE = H (High)
BACKLITE = 2 MINUTES
LIBRARY=N
Alerts: IDLE ALARM enabled
MINALERT= 10, 5 (10 Minutes, 5 Beeps)
DOSE COMPLETE enabled
VOLUME LIMIT enabled

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals