07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 7
Precautions
If the Size Override configuration option is enabled, an operator can manually
override the Syringe Recognition feature. Incorrect syringe information may cause
delivery errors.
Special Precautions for Epidural Administration
Epidural administration of anesthetics is limited to short-term infusion (not to exceed
96 hours) with indwelling catheters specifically indicated for short-term anesthetic
epidural drug delivery.
Epidural administration of analgesics is limited to use with indwelling catheters
specifically indicated for either short- or long-term analgesic epidural drug delivery.
To prevent infusion of drugs that are not indicated for epidural use, do not use IV
extension sets that incorporate injection sites during epidural delivery.
It is strongly recommended that pumps used for epidural drug delivery be clearly
differentiated from pumps used for other routes of administration.
WARNING: Epidural administration of drugs other than those indicated
for epidural use could result in serious injury to the patient.
!