07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 73
Conditions Requiring Service Cleaning
Service cleaning is a procedure for removing foreign matter from areas that are not
accessible through routine cleaning methods. Service cleaning must only be performed
by qualified individuals. Instructions for service cleaning may be found in the Auto Syringe
AS50 Infusion Pump Service Manual.
Conditions which require service cleaning include:
Spillage into the pumping mechanism, or into the barrel clamp, flange clamp, or
plunger driver mechanism.
Accumulation of foreign matter in areas not accessible to routine cleaning.
Excessive syringe-related alarms or alerts, such as CHECK SYRINGE, CHECK
PLUNGER, CHECK FLANGE
, or CHECK BARREL.
Sticking of the plunger driver, the barrel clamp, or flange clamp.