Detailed Instructions
Page 48 Auto Syringe AS50 Infusion Pump Operator’s Manual 07-19-B3-430
General Programming Information
Data Entry
The digit keys are used to enter program data.
Before pressing
CONFIRM to complete a program data entry, the value can be
changed by pressing ▲ or ▼ or can be replaced by pressing
CLR and entering
a new value. A value that appears automatically (as a result of selecting a
protocol) can be changed by pressing ▲ or ▼ or can be replaced by entering a
new value without pressing
CLR.
Field Selection
The currently selected field on the pump’s status display flashes. Use ▲ and ▼
to select the next field. Press
EDIT to open the currently selected field for
editing. When a field is opened for editing, the annunciators stop flashing and
the value itself begins flashing.
Editing in Standby State
All modes except CUSTOM DILUTION modes:
Almost all program data can be changed (edited). The exceptions are:
● Once a protocol library, protocol or infusion mode has been selected, the
only way to select a different protocol library, protocol or infusion mode is
to turn the pump off, then turn it back on and reprogram.
● The drug concentration (CONC.) and patient body weight (BODY WT) fields
cannot be changed (edited) once an infusion or bolus has been started.
To edit a programmed field:
1. Use ▲ and ▼ to select the field to be changed.
2. Press EDIT. Use ▲, ▼, and the digit keys as necessary to enter the new
value. Press
CONFIRM to complete the change.