07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 49
Detailed Instructions
CUSTOM DILUTION mode:
To make a change to syringe manufacturer, syringe size, target rate, body
weight or initial concentration before a value is entered for final volume:
1. Press CONFIRM while dashes are displayed in the FINAL VOL field.
2. Use ▲ and ▼ to select the field to be changed.
3. Press EDIT. Use ▲, ▼, and the digit keys as necessary to enter the new
value. Press
CONFIRM to complete the change.
4. Use ▲ and ▼ to return to the FINAL VOL field.
5. Press EDIT. Use ▲, ▼, and the digit keys as necessary to enter the final
volume. Press
CONFIRM to complete the entry.
6. Finish programming.
Only the final volume field will be editable after the confirm key has been
pressed and the pump starts running or delivers a bolus. To make a change to
the final volume field:
7. Press EDIT. Use ▲, ▼, and the digit keys as necessary to enter the new
value. Press
CONFIRM to complete the change.
The final concentration is displayed when a bolus or infusion is initiated.