07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 47
Detailed Instructions
Note: Pressing CONFIRM during delivery briefly displays additional program
data.
Bolus operation is not available in this delivery mode.
Example: Program the pump to infuse 5 mL over 5 minutes, using a B-D 30 mL syringe.
Stop the infusion when the TOTAL ML field shows that about 2 mL have been delivered.
Now edit the OVER field to 6 minutes (note that DOSE CANCELED appears for a few seconds),
and start the infusion.
When the pump stops after 6 minutes, about 7 mL will have been delivered; 2 mL from the
original infusion, and 5 mL from the edited infusion.