07-19-B3-430 Auto Syringe AS50 Infusion Pump Operator’s Manual Page 43
Detailed Instructions
4. Enter the dose duration (OVER) in hours and minutes. Press CONFIRM to complete
the data entry.
5. Enter the interval (EVERY) in hours and minutes. This is the time from the start of one
dose to the start of the next dose. Press
CONFIRM to complete the data entry.
6. Enter the delay (NEXT DOSE IN), then press CONFIRM. This is the time interval from
when
START is pressed until the first dose becomes due. Enter 0:00 to begin infusing
the first dose immediately.
Dose Due
Note: START must be pressed to begin infusing each dose.
The
NEXT DOSE IN field shows the time remaining until the next dose is due to be
delivered. When this time counts down to 0:00, the pump signals that the dose is due by
flashing the
ALERT and STANDBY lights, sounding continuous beeps, and showing DOSE
DUE
in the text field. START must be pressed to begin infusing each dose.
When a dose is due, pressing any key other than
START silences the audio, turns off the
DOSE DUE prompt and ALERT light, and leaves the pump in Standby state, but does not
begin dose delivery.
“Counting Time” Display
The three RUN lights flash in a “falling drop” pattern while the pump is infusing. When the
pump is counting time between scheduled doses, only the top
RUN light flashes.