10. Once the performance check is complete, click View Report.
11. Verify that the cytometer performance passed.
In the Setup tab, the cytometer performance results should have a green checkbox displayed and the word
Passed next to it.
If any parameters did not pass, see the BD Cytometer Setup and Tracking Application Guide for
troubleshooting information.
12. Select File > Exit to close the CS&T window and return to the BDFACSDiva™ interface.
The CST Mismatch dialog opens.
Click the Details button to verify which cytometer settings will be updated.
13. Click Use CST Settings.
By selecting Use CST Settings, the laser delay, area scaling, and other cytometer settings will be updated to
the latest settings from the performance check.
Chapter 6 Optimizing cytometer settings 73