Operation
Set-Up Procedures (Cont.)
00
to
11 33 88 00
(
11 77 22 55
*) (kPa)
00
to
22 00 00
(
22 55 00
*) (psig)
00
to
33 22
(in Hg)
00
to
88 00 00
(mm Hg)
*N2 and instrument air only
6. Press and release TEST button.
HIGH indicator will illuminate and
KK ,,
will be displayed. After two seconds
current high alarm setting will be
displayed. Select desired high alarm
setting by pressing either ! or "
buttons. High alarm setting ranges
are:
00
to
11 33 88 00
(
11 77 22 55
*) (kPa)
00
to
22 00 00
(
22 55 00
*) (psig)
00
to
33 22
or
NN OO NN EE
(in Hg)
00
to
88 00 00
or
NN OO NN EE
(mm Hg)
*N2 and instrument air only
7. Press and release TEST button to
save changes.
DD OO NN EE
will be
displayed for two seconds.
8. Digital display module will return to
monitoring mode.
NOTE:
If module is idle for more than one minute dur-
ing set-up procedure, module will chirp three
times, return to monitoring mode and will
default to previous settings. No changes to
settings will be saved.
The set-up must be completed and the module
must display
DD OO NN EE
in order for changes to be
saved.
Digital Display Module (Cont.)