Troubleshooting
T3300 Diagnostic Ultrasound System | 106 | B00601-010 4/10/17
*The system supports access to only one external storage device at a time. If you connect more than
one external storage device, they function in the following priority order: USB 3.0 > USB 2.0 >
microSD card.
Ethernet does not function
• The Ethernet settings are not correct.
> Contact your network administrator.
• The Ethernet cable/socket is damaged.
> Use another Ethernet cable for connection.
> Connect the system to a computer through Ethernet
connection for inspection.
Wireless connection failed
• The wireless device is turned off or not functioning.
> Check the power of the wireless device.
> Restart the wireless device.
> Connect another wireless product to this device for
inspection.
• The wireless network settings are not correct.
> Open the Quick Setup menu and check if the wireless
network function is enabled.
> Contact your network administrator.
Table 1 Troubleshooting
Symptom Possible cause and corrective action

Table of Contents