S10/S12/S10A/S12A Patient Monitor User’s Manual
11
Input impedance
5.0M
System noise
30Vpp RTI
Waveform
sweep speed
Display
6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s, error ±5%
Recorder
12.5mm/s, 25mm/s, 50mm/s, error ±5%
Waveform
Gain
Display
×0.25, ×0.5, ×1 (10mm/mV), ×2, ×4, error ±5%.
Auto
Recorder
×0.25, ×0.5, ×1 (10mm/mV), ×2, ×4, error ±5%.
CMRR
Diagnostic mode 100 dB
Monitor, Operation
mode
110 dB
Calibration voltage
±5%(×1)
Input offset current
<0.1uA
Time constant
Monitoring mode: 0.3s
Diagnostic mode: 3.2s
A.8.1.4 ECG analysis calculation
Measurement
parameter
HR, PVCs, ST, QT, and Arrhythmia analysis
Multi-lead
synchronous
analysis function
Supports synchronous analysis of at least 2 leads, one of which
is the key monitoring lead and the other is the auxiliary lead. It
is on except 3-lead mode.
Smart Lead Switch
Automatic, Manual; Default manual
(3-lead mode is fixed as manual)
Automatic mode: the algorithm automatically identifies the
current smart leads, and the host automatically switches the key
monitoring leads according to the identification of the algorithm.
HR measurement
range and accuracy
Measurement
range
Adult: 10~300 bpm
Pediatric/Neonatal: 10~350bpm。
Resolution 1 bpm
Accuracy ±1% or ±1 bpm, whichever is greater
Detection 0.20mVp-p