S12 Vet Monitor User’s Manual
8-32
Problem Corrective Actions
4. Check that clips are not dry or used for a prolonged time.
No ECG Waveform
1. Check that the ECG gain is not set too low. Adjust the gain as required.
For more information, see 8.5.5 Setting ECG Waveforms.
2. Check that the leadwires and patient cables are properly connected.
Change cable and leadwires.
3. Check that the patient cable or leadwires are not damaged. Change
them if necessary.
Base Line Wander
1. Check for excessive patient movement or muscle tremor. Secure
leadwires and cable.
2. Check that clips are not detached. For more information, see 8.4.1
Preparing the Patient Skin.
3. Check for ECG filter setting. Set ECG Filter mode to 【Monitor】.
NOTE
:
Physiological alarm and technical alarm information refer to D Alarm
Information.