S12 Vet Monitor User’s Manual
16-2
CAUTION:
In order to ensure the accuracy of dripping rate measurement, the drip monitor
module should be vertically installed or naturally hung on the infusion stand
using matching bracket.
This function is an assistive technology implementation method designed for the
high-quality infusion nursing services, and cannot replace manual monitoring
and speed control operations during infusion.
This function is suitable for working under relatively static conditions.
Therefore, avoid using it in a moving state, and avoid shaking and tilting at a
large angle. When the water mist and small water drops in the dropper are
seriously hung on the wall, it may interfere with the detection. If necessary, you
can flick the wall of the dropper with your finger to shake off the small wate
drops.
The DM module uses infrared sensing to detect, so it should be avoided in strong
light environment.
16.3. DM measurement
16.3.1. Start infusion
16.3.1.1. Connect DM module cable
Insert the DM module cable into the DM socket on the monitor, and the drip monitor
interface will be displayed on the monitor.
16.3.1.2. Pre-perfusion infusion tube
Close the flow clamp tightly and connect the infusion tube set with the infusion
container, then squeeze the drip chamber and pour the liquid medicine to the 1/2 position
of the drip chamber. Open the flow clamp, fill the liquid medicine to the tip needle, and
then close the flow clamp tightly.