S12 Vet Monitor User’s Manual
5-4
5.8. Monitor Boot Usage Configuration
When the monitor starts, you can select whether the monitor loads the nearest
configuration or the specified configuration. You can set the default configuration by
following these steps:
1. SelectMain Menuquick key→fromConfigurationcolumn to select
Boot Config→input maintenance password→Enter.
2. SelectDefaultorCurrent.
Default: The monitor loads the default configuration specified by
the user
the user when it starts. Please refer to 5.2 Selecting Default
Configuration.
Current: The monitor loads the nearest configuration when it starts.
5.9. Setting password valid time
If you access the configuration management menu and use the password to access
alarm-related settings, you can set a valid time for the password, beyond which you will
need to re-enter the password.
Please follow the steps below:
1. Select Main Menu quick key→from System column to select
Maintenance→input maintenance password→Enter.
2. SelectAuthorizationsubmenu.
3. SetRetention Time.