S12 Vet Monitor User’s Manual
23-2
If any damage or abnormal phenomenon is found, please do not use the monitor and
immediately contact the medical engineer of the hospital or the maintenance personnel of
the company.
23.2. Maintenance Schedule
The following tasks, except visual inspection, startup detection, touch screen
calibration, battery inspection and recorder inspection, can only be completed by
professional maintenance personnel. Please contact the maintenance personnel in time
when the following maintenance is required. Before testing or maintenance, the equipment
must be cleaned and disinfected.
Check / Maintenance item Recommended frequency
Preventative Maintenance Tests
Visual inspection Every day, before first use
ECG HR detect 1. When the user suspects that the measured
value is inaccurate.
2. After the relevant modules are repaired or
replaced.
3. For DM and CO2 module, at least once every
a year.
4. For other module, at least once every two
years.
Calibration detect
RESP RR detect
SpO2 SpO2 detect
PR detect
TEMP TEMP detect
NIBP Leakage detect
Static pressure detect
Pressure calibration
CO2 CO2 gas detect
IBP
Static pressure detect
Pressure calibration
DM Drip rate detect
Nurse call detect
When you suspect that the function is not normal. Analog output test
Defibrillation synchronization test
Electrical Safety Tests
Select test items based on IEC 1. After repairing or replacing the power