S12 Vet Monitor User’s Manual
13-5
13.5. Setting IBP
13.5.1. Setting the IBP alarm
You can set the alarm by following the steps below:
1. Select the IBP parameter area or waveform area to enter the IBP menu.
2. Select【Alarm】 submenu.
If the alarm setting is protected by password, enter the password. For detail,
please refer to 7.6.2 Changing Alarm Setup Protection Mode.
3. Set alarms as needed.
13.5.2. Change the pressure label
The pressure label is identifier for each type only, so the pressure label must be set up
when making pressure measurements. You can choose a pressure label following these
steps:
1. Selecting IBP parameter area or waveform area where you need to change
labels for enter corresponding IBP menu.
2. Select【Setup】submenu.
3. Select 【Label】 where the appropriate tag name in the list.
Label name Description Label name Description
PA
Pulmonary artery pressure
CVP
Central venous pressure
Ao
Aortic pressure
LAP
Left atrial pressure
UAP
Umbilical arterial pressure
RAP
Right atrial pressure
BAP
Brachial arterial pressure
ICP
Intracranial pressure
FAP
Femoral arterial pressure
UVP
Umbilical venous pressure
ART
Arterial blood pressure
LV
Left ventricular pressure
PAWP
Pulmonary wedge pressure
P1 to P2
Non-specific pressure label
NOTE:
The same label name cannot be used for different channels about IBP