15-7
Check whether the LED aperture on the module is green and always bright. (a green
light indicates that the system is normal)
Exhale toward the sampling tube and check whether CO2 waveform and value are
displayed on the monitor.
Block the sampling tube with your fingers and wait for 10s.
Check whether the blockage alarm is displayed, and whether the LED aperture on the
module flashed red at the same time.
Under appropriate circumstances, check the tightness of the patient’s breathing
circuit connected with the sampling tube.
Status indicated by LED aperture on Nomoline ISA CO
2
module:
LED aperture The indicated status
Steady green light System OK
Blinking green light Zeroing in progress
Steady red light Sensor error
Blinking red light Check sampling tube