S12 Vet Monitor User’s Manual
15-11
Sleep:When not using CO2 for monitoring, it is recommended to set the CO2
module to sleep mode to increase the life of the CO2 module.
15.7. Entering Intubation Status
When performing intubation during general anesthesia, you can enter the intubation
mode in order to reduce unnecessary alarms. To enter the intubation mode, follow this
procedure:
1. Select the CO2 parameter area or waveform area to enter theCO
2
menu.
2. SelectIntubation Statusbutton.
For the details of the intubation mode, see
7.11 Intubation Status.
15.8. Zeroing
Mainstream and Sidestream / Microflow CO
2
Module
While zeroing is recommended the first time a CO
2
sensor is connected to the monitor,
it is only absolutely necessary when the messageZero Required” is displayed.
Follow these steps:
1. Ensure that the catheter or airway adapter is not connected to the patient or
close to any source of CO
2
(including the patient's, your own exhaled breath
and ventilator exhaust valves).
2. Using any of the following methods to perform zeroing:
Select the CO2 parameter area or waveform area, and then selectZero
button.
Select Zeroquick keySelectCO
2
Zerosubmenuselect the
CO
2
to zero.
The screen promptsCO2 Zero In Progress…, and the message disappears after the
zeroing is completed.
For Nomoline ISA CO
2
, the sampling tube should be pulled out for automatic zero
when needed.