EasyManua.ls Logo

Biolight Meditech S12 Vet - Appendix A Product Specifications; Safety Specifications

Biolight Meditech S12 Vet
294 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S12 Vet Monitor User’s Manual
1
Appendix A Product Specifications
A.1 Safety Specifications
The monitor is Classified according to the IEC60601-1 is as follows:
Parts
Classifica
tion of
protectio
n against
electric
shock
Degree of
protectio
n against
electric
shock
Degree of
protectio
n against
ingress of
liquid
Degree of
protection
against
hazards of
explosion
Recommen
ded
disinfection
and
sterilization
methods
Mode of
operation
Mainframe I No mark
IP21 Not suitable
See Chapter
22
Maintenanc
e and
Cleaning of
this manual
for details.
Continuou
s
Fixed
parameter
(ECG,
TEMP,
RESP, NIBP,
SpO2)
NA
CF
IBP
C.O.
CO2
BF
DM
Note:
I: Class I, internally and externally powered equipment.
When you doubt about the protecting earth integrality or protecting earth lead of the
equipment, you’d better change the equipment to internally powered equipment.
CF: Type CF applied part.
BF: Type BF applied part.
NA: Not applicable
Not suitable: Equipment is not suitable for use in the presence of flammable
anesthetic mixture with air or with oxygen or nitrous oxide.

Table of Contents