S12 Vet Monitor User’s Manual
14-6
will calculate and display the average values of C.O. and C.I., and the average
will be stored and displayed in parameter area.
At the time of injection, the 3-way stopcock opens to the floating catheter end and
closes to the injection end. At the end of the measurement, close the end to the floating
catheter, open to the end of the injection, and then inhale the injection into the syringe.
In addition, in the C.O. measurement window, you can also choose:
Stop: Press this button during measurement to cancel this measurement.
Setup: Enter into the Setup page of the C.O. menu. This option
is displayed when the C.O. Measuring hotkey enters the C.O.
measurement window.
Hemodynamics : Enter the Calculate menu. See Chapter 19
Calculation.
NOTE:
Starting a measurement without blood temperature being stable may cause
measurement failure.
During the measurement of C.O., the blood temperature alarm is invalid. After
the measurement is finished, the blood temperature alarm will automatically
resume. Please refer to the instruction manual of the floating catheter to
determine the
Catheter Const
and the volume of the injection.
14.7. C.O. Setting
14.7.1. Setting the TB Alarm
You can set up the TB alarm by following these steps:
1. Selecting C.O. parameters area for enter the C.O. menu.
2. Select the Alarm submenu.
3. Set the alarm if needed.