S12 Vet Monitor User’s Manual
4-2
【Cancel】:Exit theoperation of discharging patient data and return to the
main interface.
4.2. Admitting a Patient
The monitor admits a new patient in the following situations:
After discharging a patient, the monitor automatically admit a new patient.
From the standby screen, select【Discharge Patient】to admit a new patient.
Always inputs patient information as soon as the patient is admitted. For more
information please refer
to 4.3.2 Editing Patient Information for details.
WARNING:
【Patient Type】 and 【Paced】 will always contain a value, regardless of
whether the patient is fully admitted or not. If you do not specify settings for
these fields, the monitor uses the default settings from the current
configuration, which might not be correct for your patient.
For paced patients, you must set 【Paced】 to 【Yes】. If it is incorrectly set to
【No】, the monitor could mistake a pace pulse for a QRS and fail to alarm
when the ECG signal is too weak.
For non-paced patients, you must set 【Paced】 to 【No】.
4.3. Managing Patient Information
4.3.1. Patient Management menu
Use any of the following methods to enter the【Patient Management】menu:
Select
patient information area at the top left corner of the screen.
Select【Patient Management】quick key.
Select 【 Main Menu 】 →from 【 Patient 】 column to select 【 Patient
Management】.