S12 Vet Monitor User’s Manual
4-2
Cancel】:Exit theoperation of discharging patient data and return to the
main interface.
4.2. Admitting a Patient
The monitor admits a new patient in the following situations:
After discharging a patient, the monitor automatically admit a new patient.
From the standby screen, selectDischarge Patientto admit a new patient.
Always inputs patient information as soon as the patient is admitted. For more
information please refer
to 4.3.2 Editing Patient Information for details.
WARNING
Patient Type and Paced will always contain a value, regardless of
whether the patient is fully admitted or not. If you do not specify settings for
these fields, the monitor uses the default settings from the current
configuration, which might not be correct for your patient.
For paced patients, you must set Paced to Yes. If it is incorrectly set to
No, the monitor could mistake a pace pulse for a QRS and fail to alarm
when the ECG signal is too weak.
For non-paced patients, you must set Paced to No.
4.3. Managing Patient Information
4.3.1. Patient Management menu
Use any of the following methods to enter thePatient Managementmenu:
Select
patient information area at the top left corner of the screen.
SelectPatient Managementquick key.
Select Main Menu →from Patient column to select Patient
Management.