S12 Vet Monitor User’s Manual
8-1
Chapter 8 ECG
8.1. Introduction
The electrocardiogram (ECG) measures the electrical activity of the heart and
displays it on the monitor as waveforms and parameters. The monitor provides 3-lead,
5-lead, 6-lead, and 12-lead ECG monitoring, arrhythmia analysis, ST segment analysis
and QT/QTc measurements.
8.2. Safety Information
WARNING
This equipment is not intended for direct cardiac application.
Make sure the conductive parts of electrodes and associated connectors for
applied parts, including the neutral electrode, should not contact any other
conductive parts including earth.
Use defibrillation-proof ECG cables during defibrillation.
Do not touch the patient or metal devices connected to the patient during
defibrillation.
To reduce the hazard of burns during high-frequency surgical procedure,
ensure that the monitor’s cables and transducers never come into contact with
the electrosurgery unit (ESU).
To reduce the hazard of burns during use of high-frequency surgical unit
(ESU), the ECG electrodes should not be located between the surgical site and
the ESU return electrode.
CAUTION
Only use parts and accessories specified in this manual. Follow the instructions
for use and adhere to all warnings and cautions.
Regularly inspect the electrode application site to ensure skin quality. If there
are signals of allergies,
replace the electrodes or change the application site.