S12 Vet Monitor User’s Manual
25
Unit
Drops/min, mL/h, can be automatically converted (for
conversion, 1mL of conventional tube =20 drops is mainly
used.)
A.8.9.2 Performance Specification
Drip rate
measurement range
5
200 Drops/min
1mL of conventional tube =20 drops
Drip accuracy
±2 and ±2% (whichever is greater)
A.9 Alarm Specification
If no special instructions are given in the following specifications, the adjustable
range of the alarm limit is the same as the measuring range of the signal.
A.9.1 ECG
Alarm limit Range Step
ST High
(low limit +0.01 Mv) ~2.00mV
0.01 mV
ST Low -2.00 mV ~ (high limit -0.01 mV)
HR High
(HR low limit +1bpm) ~400bpm
1bpm
HR Low 10bpm~ (HR high limit -1bpm)
QTc High
200ms700ms
1ms
ΔQTc Low
-500ms500ms
A.9.2 RESP
Alarm limit Range Step
RR High
(low limit +1 rpm) 150rpm
1rpm
RR Low 0rpm~ (high limit -1rpm)
A.9.3 NIBP
Alarm limit Range Step