S12 Vet Monitor User’s Manual
13-9
Set the Left Scale for arterial pressure.
Set the Right Scale for venous pressure.
Set the Grid of the overlapped waveform area.
Set the Wave Speed of the overlapped display waveform.
13.6. IBP troubleshooting
This section describes problems you may encounter during using. You can refer to
the following table for troubleshooting. If the problem persists, please contact
maintenance staff.
NOTE: See D Alarm Information for physiological alarms and technical
alarm information.
Problem Solution
IBP parameter area and waveform
area cannot be found on the
interface
1. Check whether the display of IBP
parameters is set in the Screen Layout
menu or not. For details, see 3.6.1 Setting
Parameters to be protected.
2. Check whether the IBP parameter switch is
turned on or not. For details, see 3.6.1 Setting
Parameters
3. Check the IBP cable, IBP sensor and
module are connected or not.
4. Check the valve position of the 3-way
stopcock is correct or not.
5. Confirm that the sensor has been zeroed.
For details, see 13.3.2 IBP Sensor zero.
P1/P2 does not display systolic and
diastolic pressure measurements
Set the displayed pressure to All. For details,
see 13.5.3 Setting up display types about
Extended Pressure.