S12 Vet Monitor User’s Manual
12-10
1. Select the NIBP parameter area to enter theNIBPmenu.
2. SelectSequencesubmenu.
3. Set each sequence measurement toDurationorInterval Timeseparately.
12.8.7. Setting NIBP Unit
You can change the NIBP units by following these steps:
1. Select the NIBP parameter area to enter theNIBPmenu.
2. SelectSetupsubmenu.
3. Set NIBPUnit.
12.8.8. Setting NIBP Invalid Time
NIBP measurements become outline fonts after a preset time. This avoids older NIBP
values being misinterpreted as current measurements. To set the timeout period, follow
this procedure:
1. Select Main Menu quick key→from System column to select
Maintenance→input maintenance password→Enter.
2. SelectModulesubmenu→Othersubmenu.
3. SetNIBP Invalid Time.
12.8.9. Displaying the NIBP List
To display multiple sets of the latest NIBP measurements, follow this procedure:
1. EnterScreen Layoutsubmenuby either of the following ways:
SelectScreen Setupquick key→SelectScreen Layoutsubmenu.
SelectMain Menuquick key→from Display column to select
Screen Layout.
2. In the desired parameter area, selectNIBPNIBP
List.
12.9. Assisting Venous Puncture
You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood
vessel and therefore help venous puncture. To assist venous puncture, follow this