S12 Vet Monitor User’s Manual
7-15
Intubation Status】 button.
Select【Main Menu】quick key→from【Alarm】column to select【Exit
Intubation Status】.
7.12. Other Bed Alarm
Enter other bed observation interface, and when the monitored bed monitor has an
alarm triggered, the alarm light and alarm sound are prompted according to the highest
level of all alarms of the current monitor and other bed monitor. You can view and
manage other bed
alarm.
The alarm delay time from the device to other bed is≤2s.
7.12.1. Other Bed Alarm Reset
You can reset other bed alarm on the monitor. The steps to enable it to reset the bed
alarm are as follows:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select
【Maintenance】→input maintenance password→Enter.
2. Select【Alarm】submenu→【Remote View】submenu.
3. Open【Reset Remote Bed’s Alarms】.
And then 【Bed View Settings (bed number)】 window→ 【Reset Remote Alarm】
button in the 【Alarm】 submenu will be activated. Click on 【Reset Remote Alarm】
button,
other bed alarm will be reset.
CAUTION:
Only when the “Alarm Reset By Other Bed” function of the remote monitor is
enabled, can you reset other bed alarm on this monitor
7.12.2. Authorizing the Alarm Reset to Other Devices
Alarms on your monitor can be reset by remote devices if you enable this function.
To do so, follow this procedure:
1. Select 【 Main Menu 】 quick key→from 【 System 】 column to select