3. Set【Interval Time】.
12.8.4. Selecting NIBP Start Mode
Start mode defines how NIBP auto mode works. To set the 【Start Mode】, follow
this procedure:
1. Select the NIBP parameter area to enter the【NIBP】menu
2. Select【Setup】submenu.
3. Set【Start Mode】.
【Clock】: After the first measurement, the monitor automatically
synchronizes NIBP automatic measurements with the real time clock.
For example, if 【Interval】 is set to 【30min】, and you start NIBP auto
measurement at 10:03, the next measurement will be taken at 10:30, and
then at 11:00, 11:30, and so on.
【Interval】: After the first measurement, the monitor automatically
repeats measurements at set interval. For example, if 【Interval】 is set
to【30min】, and you start NIBP auto measurement at 10:03, the next
measurement will be taken at 10:33, and then at 11:03, 11:33, and so on.
12.8.5. Enabling the NIBP End Tone
The monitor can issue a reminder tone at the completion of NIBP measurement. The
NIBP End Tone is off by default. To switch on the NIBP end tone, follow this procedure:
1. Select the NIBP parameter area to enter the【NIBP】menu
2. Select【Setup】submenu.
3. Switch on【NIBP
End Tone】.
12.8.6. Setting NIBP Sequence Measurement
NIBP sequence measurements can consist of up to 5 measurement periods: A, B, C,
D, and E. You can set the measurement duration for each period and the interval between
NIBP measurements in each period separately. The steps to set up the NIBP measurement
sequence are as follows: