S12 Vet Monitor User’s Manual
14-4
1. Selecting C.O. parameters area for enter the C.O. menu.
2. SelectSetupsubmenu.
3. Make the following settings:
Set up patient information: Enter the patient's Height and Weight,
which are used for calculate the C.I. value.
Set Catheter Const: The C.O. coefficient is a calculation coefficient
related to the floating catheter, liquid temperature measurement method
(outflow method or ice bath method), injection volume and injection
temperature. Please refer to the floating catheter manual. When replacing
a floating conduit, its coefficient should be adjusted according to
manufacturer's instructions.
Set Measure Mode. If set Continuous, the monitor automatically
performs measurement when blood temperature is stable. You do not
need to select the Start button on the C.O. measurement interface
before C.O. measurement; when setting up Single, the monitor is on,
each measurement is performed after the Start button is selected.
Set TI Source. When set Auto, the system will automatically
measure the temperature of the injection. Meanwhile, the TI Value
option is not available; when setting up Manual, you need to
manually enter the temperature of the injection at TI Value.
14.6.3. Measure C.O.
Following steps for make C.O. measurement:
1. Selecting C.O. parameters area for enter the C.O. menu.
2. Click the C.O. Measuring button at the bottom of the C.O. setup menu.