S12 Vet Monitor User’s Manual
6
A.5.7 Signal Output
Auxiliary output interface (optional)
Standard
Meet the requirements of IEC 60601-1 for short-circuit
protection and leakage current.
Output impedance Rated 50Ω
ECG analog signals output
Output signal range
-10V+10V
Maximum
transmission delay
25 ms
Sensitivity 1V/mV±5%
PACE rejection /
strengthen
Has PACE rejection function
IBP analog signals output
Output signal range
-1V+4V
Maximum
transmission delay
35 ms
Sensitivity 1V/100mHg±5%
Nurse call output
Output voltage range High level: 3.5~5V, providing a maximum of 10 mA output
current;
Low level: 0.5V, receiving a maximum of 5 mA input
current.
Isolated voltage 1500 VAC
Signal type N.C., N.O., Pulse Output (optional);
Rise and drop time ≤1ms
Defibrillator synchronization signal output
Output impedance 50Ω±10%
Maximum delay 25 ms (from R wave crest to pulse raise)
Amplitude High level: 3.5~5V, providing a maximum of 10 mA output
current;
Low level: 0.5V, receiving a maximum of 5 mA input current.