S12 Vet Monitor User’s Manual
10
A.8.1.3 ECG Measurement
Lead type
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V-
6-lead: I, II, III, aVR, aVL, aVF,Va, Vb
12-lead: I, II, III, aVR, aVL, aVF,V1~V6
Auto: identify leads automatically
Indication of lead-off shall
be provided
Every electrode
ECG abnormal work
indications
Every amplification channel shall have an indication of
abnormal ECG operation (polarization).
Bandwidth (-3dB)
Diagnostic mode: 0.05~150Hz
Monitor mode: 0.5~40Hz
Operation mode: 1~25Hz
ST mode: 0.05~40Hz
Signal quality display
Expression way: numerical display and waveform color.
Defibrillation Protection
Breakdown Voltage: 4000V 50Hz/60Hz
Anti-defibrillation effect protection: baseline recovery
time: 5s (after defibrillation).
Input signal range
Input signal range -10.0mV~+10.0Mv
Electrode offset
potential
±500 mV d.c.
Input impedance
≥5.0MΩ
System noise
≤30μVpp RTI
Waveform
sweep speed
Display
6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s, error ≤±5%
Recorder
12.5mm/s, 25mm/s, 50mm/s, error ≤±5%
Waveform
Gain
Display
×0.25, ×0.5, ×1 (10mm/mV), ×2, ×4, error ≤±5%.
Auto
Recorder
×0.25, ×0.5, ×1 (10mm/mV), ×2, ×4, error ≤±5%.
CMRR
Diagnostic mode ≥100 dB
Monitor, Operation
mode
≥110 dB