S12 Vet Monitor User’s Manual
32
Radiated RF
IEC 61000-4-3
3V/m
80MHz
2.7GHz
3V/m
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
Pd
V
1
5.3
150 KHz to 80 MHz
Pd
E
1
5.3

80 MHz
800 MHz
Pd
E
1
7

80MHz2.7GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Note 3: The device that intentionally receives RF electromagnetic energy at the exclusion band
(2400-2483.5MHz) is exempt from the ESSENTIAL PERFORMANCE requirements, but remains
safe.