S12 Vet Monitor User’s Manual
42
Item Default Setting
Low limit ART/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure: 35
mmHg
PA/PAWP: 12 mmHg
CVP/ICP/RAP/LAP/UVP venous pressure: 0 mmHg
Alarm priority MED
Alarm output OFF
CPP Alarm switch ON
High limit 90 mmHg
Low limit 30 mmHg
Alarm priority MED
Alarm output OFF
Unit ART/Ao/UAP/BAP/FAP/LV/RAP/LAP/UVP/PA/PAW
P/P1/P2: mmHg
CVP /ICP/CPP: cmH2O
Sensitivity MED
Waveform Speed 25 mm/s
Scale type Manual
Scale Upper scale ART /Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure:
160mmHg
CVP /ICP/RAP/LAP/UVP venous pressure : 20mmHg;
PA/PAWP:30mmHg
Lower scale 0mmHg
High-precision cursor
switch
OFF
High-precision cursor ART/ Ao/ UAP/BAP/FAP/LV/P1/P2 arterial pressure:
80mmHg
CVP /ICP/RAP/LAP/UVP venous pressure: 10mmHg
PA/PAWP:15mmHg
PPV measurement OFF
PPV source Auto